[Selective activation of blood pressure monitoring alarms: effect of noise pollution in the intensive care unit] - Archive ouverte HAL Access content directly
Journal Articles Annales Françaises d'Anesthésie et de Réanimation Year : 2001

[Selective activation of blood pressure monitoring alarms: effect of noise pollution in the intensive care unit]

L. Biot
  • Function : Author
L. Holzapfel
  • Function : Author
Guillaume Jean-Paul Claude Becq
C. Melot
  • Function : Author
P. Boulétreau
  • Function : Author

Abstract

OBJECTIVE: To evaluate a selective activation of sounding alarms on non-invasive blood pressure (BP) monitoring according to the patient haemodynamic status. STUDY DESIGN: Prospective study. METHODS: Activation of alarms on BP was regulated with a protocol. Sounding alarms were either inactivated when patient's haemodynamic status was stable (group 1), or activated when it was unstable (group 2). The frequency of BP measurement was one every 15 min. For all mean BP value recorded, the following criteria were analyzed: 1) normality of the value compared to ranges 65-115 mmHg in group 1 or compared to alarm thresholds in group 2; 2) consequences on the care and therapeutic; 3) delay when an abnormal value was detected and managed after more than 15 min. RESULTS: 1,674 hours of monitoring from 42 patients, allowed the analysis of 6,695 measurements of mean BP, 3,092 in group 1 and 3,603 in group 2. In group 1, 2,822 measurements were considered as normal and 3,094 measures in group 2. Eight measurements had consequences on therapeutic in group 1, with only one with delay in care giving. 287 measurements had consequences on therapeutic in group 2, 8 with delay in care giving. Six per cent of abnormal measurements in group 2 were managed with delay. This protocol reduced by 52% the production of sounding alarms on BP, without noxious effects for the patients. CONCLUSION: Selective activation of sounding alarms on BP, according to the patient haemodynamic status, reduced noise pollution and could be one solution to improve monitoring efficiency in intensive care unit.

Domains

Bioengineering
No file

Dates and versions

hal-00348270 , version 1 (18-12-2008)

Identifiers

  • HAL Id : hal-00348270 , version 1
  • PUBMED : 11695286

Cite

L. Biot, L. Holzapfel, Guillaume Jean-Paul Claude Becq, C. Melot, P. Boulétreau, et al.. [Selective activation of blood pressure monitoring alarms: effect of noise pollution in the intensive care unit]. Annales Françaises d'Anesthésie et de Réanimation, 2001, 20 (8), pp.677-85. ⟨hal-00348270⟩
1510 View
0 Download

Altmetric

Share

Gmail Facebook X LinkedIn More