Experimental validation of an ultrasonic flowmeter for unsteady flows - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Measurement Science and Technology Année : 2018

Experimental validation of an ultrasonic flowmeter for unsteady flows

Résumé

An ultrasonic flowmeter was developed for further applications in cryogenic conditions and for measuring flow rate fluctuations in the range of 0 to 70 Hz. The prototype was installed in a flow test rig, and was validated experimentally both in steady and unsteady water flow conditions. A Coriolis flowmeter was used for the calibration under steady state conditions, whereas in the unsteady case the validation was done simultaneously against two methods: particle image velocimetry (PIV), and with pressure transducers installed flush on the wall of the pipe. The results show that the developed flowmeter and the proposed methodology can accurately measure the frequency and amplitude of unsteady fluctuations in the experimental range of 0–9 l s−1 of the mean main flow rate and 0–70 Hz of the imposed disturbances
Fichier principal
Vignette du fichier
LMFL_MST_2018_DAZIN.pdf (1.52 Mo) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-02081575 , version 1 (27-03-2019)

Identifiants

Citer

Vlasios Leontidis, Christophe Cuvier, Guy Caignaert, Patrick Dupont, Olivier Roussette, et al.. Experimental validation of an ultrasonic flowmeter for unsteady flows. Measurement Science and Technology, 2018, 29 (4), pp.045303. ⟨10.1088/1361-6501/aaa65f⟩. ⟨hal-02081575⟩
56 Consultations
936 Téléchargements

Altmetric

Partager