A High Precision Fabry-Perot Cavity Polarimeter at HERA - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Journal of Instrumentation Année : 2010

A High Precision Fabry-Perot Cavity Polarimeter at HERA

Résumé

A Fabry-Perot cavity polarimeter, installed in 2003 at HERA for the second phase of its operation, is described. The cavity polarimeter was designed to measure the longitudinal polarisation of the HERA electron beam with high precision for each electron bunch spaced with a time interval of 96ns. Within the cavity the laser intensity was routinely enhanced up to a few kW from its original value of 0.7W in a stable and controllable way. By interacting such a high intensity laser beam with the HERA electron beam it is possible to measure its polarisation with a relative statistical precision of 2% per bunch per minute. Detailed systematic studies have also been performed resulting in a systematic uncertainty of 1%.

Dates et versions

in2p3-00484268 , version 1 (18-05-2010)

Identifiants

Citer

S. Baudrand, M. Bouchel, V. Brisson, R. Chiche, M. Jacquet, et al.. A High Precision Fabry-Perot Cavity Polarimeter at HERA. Journal of Instrumentation, 2010, 5, pp.P06005. ⟨10.1088/1748-0221/5/06/P06005⟩. ⟨in2p3-00484268⟩