Optical frequency comb Fourier transform cavity ring-down spectroscopy - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Optics Express Année : 2022

Optical frequency comb Fourier transform cavity ring-down spectroscopy

Résumé

We demonstrate broadband and sensitive cavity ring-down spectroscopy using a near infrared frequency comb and a time-resolved Fourier transform spectrometer. The cavity decays are measured simultaneously at each optical path difference and spectrally sorted, leading to purely exponential decays for each spectral element. The absorption spectra of atmospheric water and carbon dioxide are retrieved and demonstrate the high frequency resolution and absorption precision of the technique. The experimental apparatus, the measurement concept and the data treatment are described. The technique benefits from the advantages of cavity ring-down spectroscopy, i.e. the retrieved absorption does not depend on the cavity parameters, opening up for high accuracy absorption spectroscopy entirely calibration-free. (C) 2022 Optica Publishing Group under the terms of the Optica Open Access Publishing Agreement

Dates et versions

hal-03657659 , version 1 (03-05-2022)

Licence

Paternité

Identifiants

Citer

Romain Dubroeucq, Lucile Rutkowski. Optical frequency comb Fourier transform cavity ring-down spectroscopy. Optics Express, 2022, 30 (8), pp.13594-13602. ⟨10.1364/OE.454775⟩. ⟨hal-03657659⟩
64 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More