Calibrated work function mapping by Kelvin probe force microscopy - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Review of Scientific Instruments Année : 2018

Calibrated work function mapping by Kelvin probe force microscopy

Résumé

We propose and demonstrate the implementation of an alternative work function tip calibration procedure for Kelvin probe force microscopy under ultrahigh vacuum, using monocrystalline metallic materials with known crystallographic orientation as reference samples, instead of the often used highly oriented pyrolytic graphite calibration sample. The implementation of this protocol allows the acquisition of absolute and reproducible work function values, with an improved uncertainty with respect to unprepared highly oriented pyrolytic graphite-based protocols. The developed protocol allows the local investigation of absolute work function values over nanostructured samples and can be implemented in electronic structures and devices characterization as demonstrated over a nanostructured semiconductor sample presenting Al0.7Ga0.3As and GaAs layers with variable thickness. Additionally, using our protocol we find that the work function of annealed highly oriented pyrolytic graphite is equal to 4.6 ± 0.03 eV.
Fichier principal
Vignette du fichier
Garrillo_2018_1.5007619.pdf (3.15 Mo) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte

Dates et versions

hal-02277068 , version 1 (02-06-2022)

Identifiants

Citer

Pablo Fernández Garrillo, Benjamin Grevin, Nicolas Chevalier, Lukasz Borowik. Calibrated work function mapping by Kelvin probe force microscopy. Review of Scientific Instruments, 2018, 89 (4), pp.043702. ⟨10.1063/1.5007619⟩. ⟨hal-02277068⟩
48 Consultations
592 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More