Evaluation of a preclinical photon-counting CT prototype for pulmonary imaging

F. K. Kopp
  • Fonction : Auteur
H. Daerr
  • Fonction : Auteur
A. P. Sauter
  • Fonction : Auteur
S. Ehn
  • Fonction : Auteur
A. A. Fingerle
  • Fonction : Auteur
B. Brendel
  • Fonction : Auteur
F. Pfeiffer
  • Fonction : Auteur
E. Roessl
  • Fonction : Auteur
E. J. Rummeny
  • Fonction : Auteur
R. Proksa
  • Fonction : Auteur
P. B. Noel
  • Fonction : Auteur

Résumé

The purpose of this study was to investigate a preclinical spectral photon-counting CT (SPCCT) prototype compared to conventional CT for pulmonary imaging. A custom-made lung phantom, including nodules of different sizes and shapes, was scanned with a preclinical SPCCT and a conventional CT in standard and high-resolution (HR-CT) mode. Volume estimation was evaluated by linear regression. Shape similarity was evaluated with the Dice similarity coefficient. Spatial resolution was investigated via MTF for each imaging system. In-vivo rabbit lung images from the SPCCT system were subjectively reviewed. Evaluating the volume estimation, linear regression showed best results for the SPCCT compared to CT and HR-CT with a root mean squared error of 21.3 mm3, 28.5 mm3 and 26.4 mm3 for SPCCT, CT and HR-CT, respectively. The Dice similarity coefficient was superior for SPCCT throughout nodule shapes and all nodule sizes (mean, SPCCT: 0.90; CT: 0.85; HR-CT: 0.85). 10% MTF improved from 10.1 LP/cm for HR-CT to 21.7 LP/cm for SPCCT. Visual investigation of small pulmonary structures was superior for SPCCT in the animal study. In conclusion, the SPCCT prototype has the potential to improve the assessment of lung structures due to higher resolution compared to conventional CT.

Domaines

Imagerie

Dates et versions

hal-02078936 , version 1 (25-03-2019)

Identifiants

Citer

F. K. Kopp, H. Daerr, S. Si-Mohamed, A. P. Sauter, S. Ehn, et al.. Evaluation of a preclinical photon-counting CT prototype for pulmonary imaging. Scientific Reports, 2018, 8 (1), pp.17386. ⟨10.1038/s41598-018-35888-1⟩. ⟨hal-02078936⟩
23 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More