Z-segmentation of a transmit array head coil improves RF ramp pulse design at 7T

Abstract : In Time-Of-Flight sequences, ramp pulses such as TONE's are frequently used to compensate for thru-slab blood flow saturation in cerebral MRA. At Ultra High Field, fast-kz spokes in parallel transmission allow to mitigate B1 + heterogeneities in the slab selection process. Here we extend their use for TONE pulses and show improvement of the flip angle ramp fidelity with a homemade z-segmented head array in comparison to a purely azimuthally-distributed commercial coil.
Type de document :
Communication dans un congrès
Joint Annual Meeting ISMRM-ESMRMB 2018, Jun 2018, Paris, France. 〈https://www.ismrm.org/18m/〉
Liste complète des métadonnées

Littérature citée [11 références]  Voir  Masquer  Télécharger

https://hal.archives-ouvertes.fr/hal-01633707
Contributeur : Laurent Le Brusquet <>
Soumis le : vendredi 17 novembre 2017 - 16:50:05
Dernière modification le : mardi 27 novembre 2018 - 16:42:10
Document(s) archivé(s) le : dimanche 18 février 2018 - 12:34:42

Fichier

ismrm2018-paper1.pdf
Fichiers produits par l'(les) auteur(s)

Identifiants

  • HAL Id : hal-01633707, version 1

Citation

Gaël Saïb, Raphael Tomi-Tricot, Vincent Gras, Franck Mauconduit, Nicolas Boulant, et al.. Z-segmentation of a transmit array head coil improves RF ramp pulse design at 7T. Joint Annual Meeting ISMRM-ESMRMB 2018, Jun 2018, Paris, France. 〈https://www.ismrm.org/18m/〉. 〈hal-01633707〉