Abdominal aortic aneurysm imaging with 3-D ultrasound: 3-D-based maximum diameter measurement and volume quantification. - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Ultrasound in Medicine & Biology Année : 2013

Abdominal aortic aneurysm imaging with 3-D ultrasound: 3-D-based maximum diameter measurement and volume quantification.

Résumé

The clinical reliability of 3-D ultrasound imaging (3-DUS) in quantification of abdominal aortic aneurysm (AAA) was evaluated. B-mode and 3-DUS images of AAAs were acquired for 42 patients. AAAs were segmented. A 3-D-based maximum diameter (Max3-D) and partial volume (Vol30) were defined and quantified. Comparisons between 2-D (Max2-D) and 3-D diameters and between orthogonal acquisitions were performed. Intra- and inter-observer reproducibility was evaluated. Intra- and inter-observer coefficients of repeatability (CRs) were less than 5.18 mm for Max3-D. Intra-observer and inter-observer CRs were respectively less than 6.16 and 8.71 mL for Vol30. The mean of normalized errors of Vol30 was around 7%. Correlation between Max2-D and Max3-D was 0.988 (p < 0.0001). Max3-D and Vol30 were not influenced by a probe rotation of 90°. Use of 3-DUS to quantify AAA is a new approach in clinical practice. The present study proposed and evaluated dedicated parameters. Their reproducibility makes the technique clinically reliable.

Dates et versions

hal-00921670 , version 1 (20-12-2013)

Identifiants

Citer

A. Long, L. Rouet, A. Debreuve, R. Ardon, C. Barbe, et al.. Abdominal aortic aneurysm imaging with 3-D ultrasound: 3-D-based maximum diameter measurement and volume quantification.. Ultrasound in Medicine & Biology, 2013, 39 (8), pp.1325-36. ⟨10.1016/j.ultrasmedbio.2013.03.008⟩. ⟨hal-00921670⟩
49 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More