Registration-based geometric calibration of industrial X-ray tomography system

Abstract : This paper presents a general and efficient method for geometrical calibration of industrial X-ray tomography systems without the use of any phantom. In the field of nondestructive testing (NDT), an accurate numericalmodel of the object is usually available. By extending the classical volume-to-image registration framework, the method estimates both the parameters related to the position of the object, and those related to the positions and the orientations of the X-ray source, the detector and the rotation axis. Only using the projections data and the numericalmodel of the object, the values of the geometrical parameters are estimated with a good accuracy. Numerical experiments show that the method provides good results, even if the initialization of the process is far from the real values.
Document type :
Journal articles
Complete list of metadatas

https://hal.archives-ouvertes.fr/hal-00878581
Contributor : Valerie Kaftandjian <>
Submitted on : Wednesday, October 30, 2013 - 12:24:15 PM
Last modification on : Monday, February 10, 2020 - 6:12:33 PM

Identifiers

Collections

Citation

Ismail Ben Tekaya, Valerie Kaftandjian, Fanny Buyens, Sylvie Sevestre-Ghalila, Samuel Legoupil. Registration-based geometric calibration of industrial X-ray tomography system. IEEE Transactions on Nuclear Science, Institute of Electrical and Electronics Engineers, 2013, 60, pp.1-8. ⟨10.1109/TNS.2013.2279675⟩. ⟨hal-00878581⟩