Article Dans Une Revue Review of Scientific Instruments Année : 2012

In-situ comprehensive calibration of a tri-port nano-electro-mechanical device

Résumé

We report on experiments performed in vacuum and at cryogenic temperatures on a tri-port nano-electro-mechanical (NEMS) device. One port is a very non-linear capacitive actuation, while the two others implement the magnetomotive scheme with a linear input force port and a (quasilinear) output velocity port. We present an experimental method enabling a full characterization of the nanomechanical device harmonic response: the non-linear capacitance function C(x) is derived, and the normal parameters k and m (spring constant and mass) of the mode under study are measured through a careful definition of the motion (in meters) and of the applied forces (in Newtons). These results are obtained with a series of purely electric measurements performed without disconnecting/reconnecting the device, and rely only on known DC properties of the circuit, making use of a thermometric property of the oscillator itself: we use the Young modulus of the coating metal as a thermometer, and the resistivity for Joule heating. The setup requires only three connecting lines without any particular matching, enabling the preservation of a high impedance NEMS environment even at MHz frequencies. The experimental data are fit to a detailed electrical and thermal model of the NEMS device, demonstrating a complete understanding of its dynamics. These methods are quite general and can be adapted (as a whole, or in parts) to a large variety of elecromechanical devices
Fichier principal
Vignette du fichier
Collin_v4.pdf (734.48 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-00726002 , version 1 (17-12-2013)

Identifiants

Citer

Eddy Collin, Martial Defoort, Kunal Lulla, Thomas Moutonet, Jean-Savin Heron, et al.. In-situ comprehensive calibration of a tri-port nano-electro-mechanical device. Review of Scientific Instruments, 2012, 83 (4), pp.045005. ⟨10.1063/1.4705992⟩. ⟨hal-00726002⟩

Collections

UGA CNRS NEEL
108 Consultations
131 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More