Evaluation of a purge-and-trap injection system for capillary gas chromatography-microwave induced plasma-atomic emission spectrometry for the determination of volatile selenium compounds in water - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Mikrochimica Acta Année : 1995

Evaluation of a purge-and-trap injection system for capillary gas chromatography-microwave induced plasma-atomic emission spectrometry for the determination of volatile selenium compounds in water

M.B. de La Calle Guntinas
  • Fonction : Auteur
M. Ceulemans
  • Fonction : Auteur
C. Witte
  • Fonction : Auteur
F.C. Adams
  • Fonction : Auteur

Résumé

A method is presented for the selective determination of the volatile selenium species dimethylselenide and dimethyldiselenide, using a commercially available purge-and-trap injection system coupled to capillary gas chromatography-microwave induced plasma-atomic emission spectrometry. The efficiency of the purging step was evaluated and the parameters affecting the purge and trap processes were optimized. The method was applied to the determination of volatile selenium compounds in lake water. Relative detection limits of 2ng/l for dimethylselenide and dimethyldiselenide, corresponding to an absolute detection limit of 10 pg, were achieved. © 1995 Springer-Verlag.

Dates et versions

hal-00291989 , version 1 (30-06-2008)

Identifiants

Citer

M.B. de La Calle Guntinas, M. Ceulemans, C. Witte, Ryszard Lobinski, F.C. Adams. Evaluation of a purge-and-trap injection system for capillary gas chromatography-microwave induced plasma-atomic emission spectrometry for the determination of volatile selenium compounds in water. Mikrochimica Acta, 1995, 120 (1-4), pp.73-82. ⟨10.1007/BF01244421⟩. ⟨hal-00291989⟩