High-precision efficiency calibration of a high-purity co-axial germanium detector

Bertram Blank
J. Souin
  • Fonction : Auteur
P. Ascher
L. Audirac
  • Fonction : Auteur
G. Canchel
  • Fonction : Auteur
M. Gerbaux
S. Grevy
  • Fonction : Auteur
J. Giovinazzo
H. Guerin
  • Fonction : Auteur
T. Kurtukian Nieto
I. Matea
  • Fonction : Auteur

Résumé

A high-purity co-axial germanium detector has been calibrated in efficiency to a precision of about 0.15% over a wide energy range. High-precision scans of the detector crystal and gamma-ray source measurements have been compared to Monte-Carlo simulations to adjust the dimensions of a detector model. For this purpose, standard calibration sources and short-lived on-line sources have been used. The resulting efficiency calibration reaches the precision needed e.g. for branching ratio measurements of super-allowed beta decays for tests of the weak-interaction standard model.

Dates et versions

in2p3-00975582 , version 1 (08-04-2014)

Identifiants

Citer

Bertram Blank, J. Souin, P. Ascher, L. Audirac, G. Canchel, et al.. High-precision efficiency calibration of a high-purity co-axial germanium detector. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2015, 776, pp.34-44. ⟨10.1016/j.nima.2014.12.071⟩. ⟨in2p3-00975582⟩
66 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More