Response functions of imaging plates to photons, electrons and 4He particles - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Review of Scientific Instruments Année : 2013

Response functions of imaging plates to photons, electrons and 4He particles

Résumé

Imaging plates from Fuji (BAS-SR, MS, and TR types) are phosphor films routinely used in ultra high intensity laser experiments. However, few data are available on the absolute IP response functions to ionizing particles. We have previously measured and modeled the IP response functions to protons. We focus here on the determination of the responses to photons, electrons, and 4He particles. The response functions are obtained on an energy range going from a few tens of keV to a few tens of MeV and are compared to available data. The IP sensitivities to the different ionizing particles demonstrate a quenching effect depending on the particle stopping power.

Dates et versions

in2p3-00936577 , version 1 (27-01-2014)

Identifiants

Citer

T. Bonnet, M. Comet, D. Denis-Petit, F. Gobet, F. Hannachi, et al.. Response functions of imaging plates to photons, electrons and 4He particles. Review of Scientific Instruments, 2013, 84, pp.103510. ⟨10.1063/1.4826084⟩. ⟨in2p3-00936577⟩