BACKGROUND: Estrogens are involved in the natural history of the prostate cancer and estrone sulfate, the quantitatively main circulating plasma estrogen in men, has been associated with an aggressive form of this cancer. A convenient and accurate plasma assay of this steroid has become important. METHODS: We simultaneously assayed estrone sulfate in the plasma of one hundred men aged 30-50 years, according to LC-MS/MS, GC-MS after solvolysis of E(1)S, radioimmunoassay after a chromatographic purification step, and a direct RIA commercial kit. RESULTS: Estrone sulfate plasma levels obtained with the first three methods were not significantly different. However, estrone sulfate levels measured by the direct RIA were three-fold higher than those obtained by the first three methods. We showed that the excessively high estrone sulfate levels obtained with the direct RIA kit had two origins: interference by high dehydroepiandrosterone sulfate plasma levels in men, and estrone sulfate inaccurate low concentrations in the standards. CONCLUSION: The LC-MS/MS method can be considered as an optimum option for clinical laboratory. The GC-MS method requires solvolysis to estrone, but allows simultaneous unconjugated steroid measurement. RIA method, with chromatographic purification, is cumbersome, but less expensive. DSL-5400 kit yielded estrone sulfate plasma levels that were too high.
Fichier principal
Vignette du fichier
Manuscript_CCA_Plasma_estrone_sulfate_assay_in_men_.pdf (393.4 Ko) Télécharger le fichier
Origin : Files produced by the author(s)
Loading...

Dates and versions

inserm-00515188 , version 1 (06-09-2010)

Identifiers

Cite

Frank Giton, Patrick Caron, René Bérubé, Alain Bélanger, Olivier Barbier, et al.. Plasma estrone sulfate assay in men: Comparison of radioimmunoassay, mass spectrometry coupled to gas chromatography (GC-MS), and liquid chromatography-tandem mass spectrometry (LC-MS/MS).: Plasma estrone sulfate assay in men: comparison between four different assay methods. Clinica Chimica Acta, 2010, 411 (17-18), pp.1208-13. ⟨10.1016/j.cca.2010.04.022⟩. ⟨inserm-00515188⟩

Collections

INSERM IMRB UPEC
373 View
809 Download

Altmetric

Share

Gmail Facebook X LinkedIn More