Impact of additional mattresses in emergency CT on the automated patient centering proposed by a 3D camera: a phantom study

Résumé

Abstract To assess the impact of the use of additional mattresses of different thicknesses on radiation dose and image noise based on the patient centering proposed by a 3D camera for CT. An anthropomorphic phantom was placed on mattresses of different thicknesses (from 3.5 to 13.5 cm) on the table of a CT scanner. The automated patient centering proposed by a 3D camera was analysed as a function of mattress thickness and corrected for table height. For this purpose, the impact on image noise in the lung tissues in the chest area and in the soft tissues in the abdomen-pelvis area, modulated mAs (mAs mod ) by the tube current modulation system (TCM) and volume CT dose index (CTDI vol ) was assessed slice-by-slice along the z-axis after CT scans. With the use of a mattress, the automated centering proposed by the 3D camera resulted in placement of the phantom above the isocentre. This incorrect positioning led to a significant increase in the mAs mod along the z-axis (p < 0.05) and in the CTDI vol . Image noise was significantly higher (p < 0.05) for automated phantom centering than with manual phantom centering. Differences of image noise between acquisitions with mattresses after automatic and manual phantom centering increased with the mattress thicknesses. The use of an additional mattress placed between the patient’s back and the table-top would require correcting the vertical centering proposed by the 3D camera. This manual correction is essential to avoid increased dose delivered to the patient and higher image noise.

Domaines

Imagerie
Fichier principal
Vignette du fichier
s41598-021-92637-7.pdf (2.05 Mo) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte

Dates et versions

hal-03633922 , version 1 (07-04-2022)

Licence

Paternité

Identifiants

Citer

Joël Greffier, Julien Frandon, Hélène de Forges, Aymeric Hamard, Asmaa Belaouni, et al.. Impact of additional mattresses in emergency CT on the automated patient centering proposed by a 3D camera: a phantom study. Scientific Reports, 2021, 11, pp.13191. ⟨10.1038/s41598-021-92637-7⟩. ⟨hal-03633922⟩
18 Consultations
30 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More