Design and in vitro studies of a needle-type glucose sensor for subcutaneous monitoring - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Analytical and Bioanalytical Chemistry Année : 1991

Design and in vitro studies of a needle-type glucose sensor for subcutaneous monitoring

Résumé

A new miniaturized glucose oxidase based needle-type glu¬ cose mlcrosensor has been developed for subcutaneous glu¬ cose monitoring. The sensor Is equivalent In shape and size to a 26-gauge needle (0.45-mm o.d.) and can be Implanted with ease without any Incision. The novel configuration greatly facilitates the deposition of enzyme and polymer films so that sensors with characteristics suitable for In vivo use (upper limit of linear range > 15 mM, response time < 5 mln, and sensitivity yielding a 5:1 signal-to-background ratio at normal basal glucose levels) can be prepared In high yield (>60%). The sensor response is largely Independent of ox¬ ygen tension In the normal physiological range. It also ex¬ hibits good selectivity against common interferences except for the exogenous drug acetaminophen.
Fichier principal
Vignette du fichier
ENPC00_AR_LEESU_0040.pdf (16.55 Mo) Télécharger le fichier
Origine : Accord explicite pour ce dépôt
Loading...

Dates et versions

hal-01179862 , version 1 (23-07-2015)

Identifiants

Citer

Dilbir S. Bindra, Yanan Zhang, George S. Wilson, Robert Sternberg, Thévenot Daniel R., et al.. Design and in vitro studies of a needle-type glucose sensor for subcutaneous monitoring. Analytical and Bioanalytical Chemistry, 1991, 63, pp.1692-1696. ⟨10.1021/ac00017a008⟩. ⟨hal-01179862⟩
209 Consultations
1636 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More