A New Configuration of Multipolar Cuff Electrode and Dedicated IC for Afferent Signal Recording

Serge Bernard 1 Yves Bertrand 1 Guy Cathébras 2 Lionel Gouyet 2 David Guiraud 3, 2
1 SysMIC - Conception et Test de Systèmes MICroélectroniques
LIRMM - Laboratoire d'Informatique de Robotique et de Microélectronique de Montpellier
2 DEMAR - Artificial movement and gait restoration
CRISAM - Inria Sophia Antipolis - Méditerranée , LIRMM - Laboratoire d'Informatique de Robotique et de Microélectronique de Montpellier
Abstract : Sensory information coming from natural sensors and being propagated on afferent nerve fibers could be used as feedback for a more efficient closed-loop control of a Functional Electrical Stimulation system. In order to extract and separate these signals according to their nerve fascicule origins, we propose a new architecture of a multipolar cuff electrode and an optimized integrated acquisition circuit. Concerning the electrode, we propose a specific configuration using a large number of poles in order to both reject parasitic signals, such as electromyogram and provide a maximum of recording channels in order to help the signal localization inside the nerve. Moreover, specific low-level analog signal processing was designed to extract the expected low-amplitude signal from its noisy environment. This signal processing is implemented in an ASIC that has to be implanted close to the electrode to achieve the best signal-to-noise ratio.
Complete list of metadatas

Cited literature [13 references]  Display  Hide  Download

https://hal-lirmm.ccsd.cnrs.fr/lirmm-00195236
Contributor : Serge Bernard <>
Submitted on : Monday, December 10, 2007 - 11:45:38 AM
Last modification on : Thursday, August 29, 2019 - 1:11:19 AM
Long-term archiving on: : Monday, April 12, 2010 - 6:44:11 AM

Identifiers

  • HAL Id : lirmm-00195236, version 1

Collections

Citation

Serge Bernard, Yves Bertrand, Guy Cathébras, Lionel Gouyet, David Guiraud. A New Configuration of Multipolar Cuff Electrode and Dedicated IC for Afferent Signal Recording. 3rd International IEEE/EMBS Conference on Neural Engineering, Hawai, United States. pp.578-581. ⟨lirmm-00195236⟩

Share

Metrics

Record views

527

Files downloads

387