Impact of four kVp combinations available in a dual‐source CT on the spectral performance of abdominal imaging: A task‐based image quality assessment on phantom data - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Journal of Applied Clinical Medical Physics Année : 2021

Impact of four kVp combinations available in a dual‐source CT on the spectral performance of abdominal imaging: A task‐based image quality assessment on phantom data

Résumé

Purpose: To compare the spectral performance of four combinations of kVp available in a third generation dual-source CT (DSCT) on abdominal imaging. Methods: An image-quality phantom was scanned with a DSCT using four kVp pairs (tube "A" voltage/tube "B" voltage): 100/Sn150 kVp, 90/Sn150 kVp, 80/Sn150 kVp, and 70/Sn150 kVp, classic parameters and dose level for abdomen examination (CTDIvol : 11 mGy). The noise power spectrum (NPS) and the task-based transfer function (TTF) of two inserts were computed on virtual monochromatic images (VMIs) at 40/50/60/70 keV and for mixed, low-, and high-kVp images. Detectability index (d') was computed on VMIs and mixed images to model the detection task of liver metastasis (LM) and hepatocellular carcinoma (HCC). Iodine quantification accuracy was assessed using the Root Mean Square Deviation (RMSDiodine ) and the iodine bias (IB). Results: Noise magnitude decreased by -55%± 0% between 40 and 70 keV for all kVp pairs. Compared to 70/Sn150 kVp, noise magnitude was increased by 9% ± 0% with 80/Sn150 kVp, by 16% ± 1% with 90/Sn150 kVp and by 24%± 1% with 100/Sn150 kVp. The average NPS spatial frequency (fav ) shifted toward higher frequencies as energy level increased for all kVp pairs. Lowest fav values were found for 70/Sn150 kVp and highest for 100/Sn150 kVp. The value of TTF at 50% (f50 ) shifted toward lower frequencies with increasing energy level. The highest f50 values occurred for 100/Sn150 kVp and the lowest for 80/Sn150 kVp. For both lesions, d' was highest for 70/Sn150 kVp and lowest for 100/Sn150 kVp. Compared to 70/Sn150 kVp, d' decreased by -6% ± 3% with 80/Sn150 kVp, by -11% ± 2% with 90/Sn150 kVp and by -13%± 2% with 100/Sn150 kVp. For all acquisitions, the RSMDiodine and IB were the lowest for 100/Sn150 kVp (0.29 ± 0.10 mg/ml and 0.88 ± 0.30 mg/ml, respectively) and increased when the tube "A" voltage decreased (2.34 ± 0.29 mg/ml for 70/Sn150 kVp and 7.42 ± 0.51 mg/ml respectively). Conclusion: 70/Sn150 kVp presented the lowest image noise and highest detectability in VMIs of two small focal liver lesions. 100/Sn150 kVp presented the lowest image noise on mixed images and highest accuracy of iodine quantification in iodine images.
Fichier principal
Vignette du fichier
J Applied Clin Med Phys - 2021 - Greffier - Impact of four kVp combinations available in a dual%u2010source CT on the spectral.pdf (1.65 Mo) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte

Dates et versions

hal-03631384 , version 1 (05-04-2022)

Licence

Paternité

Identifiants

Citer

Joël Greffier, Julien Frandon, Alexandre Sadate, Philippe Akessoul, Asmaa Belaouni, et al.. Impact of four kVp combinations available in a dual‐source CT on the spectral performance of abdominal imaging: A task‐based image quality assessment on phantom data. Journal of Applied Clinical Medical Physics, 2021, 22 (8), pp.243-254. ⟨10.1002/acm2.13369⟩. ⟨hal-03631384⟩
26 Consultations
18 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More