Evaluation of a new point-of-care testing for creatinine and urea measurement - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Scandinavian Journal of Clinical and Laboratory Investigation Année : 2021

Evaluation of a new point-of-care testing for creatinine and urea measurement

Résumé

Point of care testing makes it possible to obtain results in an extremely short time. Recently, radiometer has expanded the panel of tests available on its ABL90 FLEX PLUS blood gas analyzer (ABL90) by adding urea and creatinine. The aim of this study was to verify the performance of these new parameters. This included assessment of imprecision, linearity, accuracy by comparison with central laboratory standard assays and interferences. In addition, clinical utility in a dialysis center was evaluated. Within-lab coefficients of variation were close to 2%. The mean and limits of agreement (mean ± 1.96 SD) of the difference between ABL90 and Roche enzymatic assays on cobas 8000 were 0.5 (from -1.4 to 2.3) mmol/L and -0.9 (from -19.5 to 17.8) µmol/L for urea and creatinine, respectively. The ABL90 enzymatic urea and creatinine assays met the acceptance criteria based on biological variation for imprecision and showed good agreement with central laboratory. The two assays were unaffected by hematocrit variation between 20 and 70%, hemolysis and icterus interferences. It should be noted that the relationship between lab methods and ABL90 was conserved even for high pre-dialysis values allowing easy access to dialysis adequacy parameters (Kt/V) and muscle mass evaluation (creatinine index). Rapid measurement of creatinine and urea using whole blood specimens on ABL90 appears as a fast and convenient method. Analytical performances were in accordance with our expectations without any significant interferences by hemolysis or icterus.
Fichier principal
Vignette du fichier
2021 Bargnoux et al., evaluation.pdf (28.77 Mo) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-03211936 , version 1 (29-04-2021)

Identifiants

Citer

Anne-Sophie Bargnoux, Nils Kuster, Thibault Sutra, Laëtitia Laroche, Annie Rodriguez, et al.. Evaluation of a new point-of-care testing for creatinine and urea measurement. Scandinavian Journal of Clinical and Laboratory Investigation, In press, 81 (4), pp.280-297. ⟨10.1080/00365513.2021.1914344⟩. ⟨hal-03211936⟩
47 Consultations
91 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More