Skip to Main content Skip to Navigation
Conference papers

REAL-TIME MONITORING OF AIRBORNE FORMALDEHYDE CONCENTRATIONS BASED ON A MICROFLUIDIC ANALYTICAL DEVICE: APPLICATION TO FIELD MEASUREMENTS

Abstract : A new analytical method based on microfluidic device for formaldehyde detection was developed and patented. This 4 kg fully automatic instrument has a response time of 10 min, a temporal resolution of 2 seconds and a detection limit of 1 µg m-3. Our formaldehyde micro-analyser was then used for continuous monitoring during field campaigns in schools for several weeks. Our measurements were found to be in excellent agreement with those measured with the ISO 16000-3 reference method (active sampling on DNPH cartridges).
Complete list of metadatas

Cited literature [3 references]  Display  Hide  Download

https://hal.archives-ouvertes.fr/hal-02454398
Contributor : Stéphane Le Calvé <>
Submitted on : Friday, January 24, 2020 - 2:27:43 PM
Last modification on : Tuesday, May 19, 2020 - 1:11:51 AM
Long-term archiving on: : Saturday, April 25, 2020 - 3:24:52 PM

File

PG0926.pdf
Files produced by the author(s)

Identifiers

  • HAL Id : hal-02454398, version 1

Collections

Citation

Stéphane Le Calvé, Claire Trocquet, Pierre Bernhardt, Maud Guglielmino, Christina Andrikopolou. REAL-TIME MONITORING OF AIRBORNE FORMALDEHYDE CONCENTRATIONS BASED ON A MICROFLUIDIC ANALYTICAL DEVICE: APPLICATION TO FIELD MEASUREMENTS. µTAS 2018, Nov 2020, Kaohsiung, Taiwan. ⟨hal-02454398⟩