Do Bevacizumab solutions interact during an infusion through implantable venous access ports with silicone or polyurethane catheters ?

Résumé

This work aims to evaluate the possible impact of interactions between bevacizumab solutions and an implantable port equipped with a silicone or a polyurethane catheter after infusion through a complete infusion setup in simulated use conditions. Physicochemical and structural stability of bevacizumab solution was assessed by: visual examination, subvisible particles counting, dynamic light scattering, size exclusion chromatography and ion exchange chromatography. Mechanical properties of the catheters were evaluated by measuring Shore-A hardness, strain at break, strain at stress and Young's modulus. The physicochemical surface state of the catheters was assessed by FTIR-ATR spectroscopy, scanning electron microscopy (SEM) and by water contact angle measurement. Analysis of the bevacizumab solution did not highlight any signs of instability or loss of active substance. Mechanical properties of both materials remained unchanged after the infusion. During material analysis, a decrease of water contact angle observed after infusion and was more pronounced for polyurethane catheters than for silicone's, possibly due to bevacizumab adsorption or possible leachable extraction from the materials. Surface modifications were also noticed by SEM. This study did not highlight any modifications that could alter the quality of the bevacizumab infusion, nor of the infusion catheter in polyurethane or silicone, despite a modification of surface hydrophilicity. Even if after a single infusion implantable ports remained safe to use, they aim to be used for several infusion of various drugs during its lifetime, and further studies are needed to assess the impact of repeated infusions.
Fichier principal
Vignette du fichier
2019 - bevacizumab - Post print.pdf (1.83 Mo) Télécharger le fichier
2019 - bevacizumab - Supplementary data.pdf (337.34 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-02385938 , version 1 (29-11-2019)

Identifiants

Citer

Nicolas Tokhadze, Philip Chennell, Régis Cueff, Valérie Sautou. Do Bevacizumab solutions interact during an infusion through implantable venous access ports with silicone or polyurethane catheters ?. Journal of the Royal Society Interface, 2019, 16 (158), pp.20180721. ⟨10.1098/RSIF.2018.0721⟩. ⟨hal-02385938⟩
131 Consultations
126 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More