Simultaneous assay of ascorbate and urate antioxidants in human blood serum using PEDOT-based electrocehmical microsensor

Abstract : An electrochemical microsensor has been developed for the simultaneous assay of ascorbate (AA) and urate (UA) antioxidants in human blood serum. The electrode surface was modified by means of electropolymerized conductive poly(3,4-ethylenedioxythiophene) PEDOT organic films. Highly selective responses (detection potential difference more than 250 mV) were obtained for both biomarkers by optimizing the PEDOT thickness. Contrary to most previous studies the analytical performances were recorded by testing the microsensor directly in blood serum without any pretreatment of the samples. Using differential pulse voltammetry (DPV) the sensitivity and detection limit were 0.481 µA µM-1 cm-2 and 4.2 µmol L-1 for AA and 1.815 µA µM-1 cm-2 and 2.0 µmol L-1 for UA. The calibration curves were linear in the concentration range 5–200 µmol L-1 for AA and 3–700 µmol L-1 for UA. The accuracy of the sensor was satisfactorily compared with the reference analytical methods provided that the synergic effect between both antioxidants was considered. The sensor response was unmodified in the presence of the major biochemical interfering species.
Complete list of metadatas

https://hal.archives-ouvertes.fr/hal-02355061
Contributor : Open Archive Toulouse Archive Ouverte (oatao) <>
Submitted on : Friday, November 8, 2019 - 9:37:51 AM
Last modification on : Saturday, November 9, 2019 - 1:33:06 AM

File

Seklibelaidi_8948.pdf
Files produced by the author(s)

Identifiers

Collections

Citation

Fadhila Sekli-Belaidi, Anne Galinier, Pierre Gros. Simultaneous assay of ascorbate and urate antioxidants in human blood serum using PEDOT-based electrocehmical microsensor. Combinatorial Chemistry & High Throughput Screening, 2013, 16 (2), pp.84-91. ⟨10.2174/1386207311316020002⟩. ⟨hal-02355061⟩

Share

Metrics

Record views

1

Files downloads

6