A Signal Processing Method for Artefact Rejection in Transcranial Doppler Signals used for Micro-embolus detection

Résumé

Artefacts are High Intensity Transient Signals that appear in the Doppler signal due to probe tapping, probe displacement, patient movement and other external factors during TCD recording. In Holter TCD, the number of artefacts are further increased due to the servo-controlled probe positioning and long recording time. Current artefact rejection methods must be adapted to the new holter devices. Therefore, in this paper we present a complete procedure for eliminating two types of artefacts that occur during Holter TCD. The latter two types are bidirectional areftact, occurring in the energy of both positive and negative frequencies, and unidirectional artefacts occuring only in the energy of the negative frequencies. From a dataset of 25 signals, 218 artefact signatures were identified; 95% of which are bidirectional and 5% unidirectional. As a final result, 98% of the artefacts where successfully removed.
Fichier principal
Vignette du fichier
1570560411.pdf (1005.6 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-02346758 , version 1 (05-11-2019)

Identifiants

Citer

Maroun Geryes, Sébastien Ménigot, Jamal Charara, Ahmad Skaiky, Ali Mcheick, et al.. A Signal Processing Method for Artefact Rejection in Transcranial Doppler Signals used for Micro-embolus detection. ICABME 2019, Oct 2019, Tripoli, Lebanon. ⟨10.1109/ICABME47164.2019.8940150⟩. ⟨hal-02346758⟩
77 Consultations
103 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More