Comparison of electromagnetic transmitter and ultrasound imaging for intrafraction monitoring of prostate radiotherapy

Résumé

Background and purpose: To compare two in-beam monitoring devices for prostate radiotherapy: intra-prostatic electromagnetic transmitters (EM-T) (RayPilot®, Micropos Medical) and ultrasound imaging using transperineal probe (TP-US) (Clarity®, Elekta) used concomitantly on phantom and on patients. Materials and methods: The phantom study evaluated accuracy in presence of known translations and rotations. Then intra-fraction motions were analyzed for 10 prostate cancer patients implanted with the EM-T 8 days before the simulation CT (171 sessions). The percent time in which the differences between the systems were 1–5 mm were scored for each direction. Results: Experiments on phantom confirmed no interference between the systems and showed deviations of less than 0.5 mm when translations were applied progressively. In presence of rotations (5–15°), both systems displayed systematic shifts up to 6.9 and 3.8 mm for the TP-US and the EM-T, respectively. Absolute mean differences between displacements observed on patients with EM-T and TP-US were ≤0.55 mm in all directions except for one patient (≤1.77 mm). With an exception for this patient, a strong correlation was found in left–right direction: differences >2 mm were monitored less than 0.22% of the time (mean acquisition time:164 minutes) and never exceeded 5 s. Maximum differences were observed in supero-inferior direction with differences >2 mm monitored more than 6.5% of the time for 3 patients. Large prostate rotations, the presence of gas and EM-T location in the prostate may explain important differences. Conclusion: Apart from the systematic shifts induced by the rotations, the two systems were correlated and represent feasible solutions for monitoring prostate cancer treatment.
Fichier principal
Vignette du fichier
biston2019.pdf (2.09 Mo) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-02310928 , version 1 (05-10-2020)

Identifiants

Citer

Marie-Claude Biston, Timothée Zaragori, Laurent Delcoudert, Marie Fargier-Voiron, Alexandre Munoz, et al.. Comparison of electromagnetic transmitter and ultrasound imaging for intrafraction monitoring of prostate radiotherapy. Radiotherapy & Oncology, 2019, 136, pp.1-8. ⟨10.1016/j.radonc.2019.03.020⟩. ⟨hal-02310928⟩
101 Consultations
168 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More