Electromagnetic Characteristics Measurement Setup at Variable Temperatures Using a Coaxial Cell - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Advances in Materials Science and Engineering Année : 2019

Electromagnetic Characteristics Measurement Setup at Variable Temperatures Using a Coaxial Cell

Thibaut Letertre
  • Fonction : Auteur
Philippe Pouliguen
Pierre Sabouroux

Résumé

is article presents an easy way to measure frequency dependence of the complex permittivity and complex permeability for any kind of material from 10 MHz to 6 GHz, with temperature variation between the ambient temperature and 85°C. is work is based on a well-known transmission/reflection technique using a coaxial line equipped with a thermoregulation system to manage temperature variations in the sample confinement area. e paper underlines some effects which have to be taken into account with temperature variation. APU10 and cyclohexanol are presented as examples of solid and liquid reference materials.
Fichier principal
Vignette du fichier
ARTICLE-3646979.pdf (2.68 Mo) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte
Loading...

Dates et versions

hal-02297140 , version 1 (25-09-2019)

Identifiants

Citer

Thibaut Letertre, Philippe Pouliguen, Pierre Sabouroux. Electromagnetic Characteristics Measurement Setup at Variable Temperatures Using a Coaxial Cell. Advances in Materials Science and Engineering, 2019, 2019, pp.3646979. ⟨10.1155/2019/3646979⟩. ⟨hal-02297140⟩