Silicon Photomultiplier for Medical Imaging -Analysis of SiPM characteristics- - Archive ouverte HAL Accéder directement au contenu
Pré-Publication, Document De Travail Année : 2019

Silicon Photomultiplier for Medical Imaging -Analysis of SiPM characteristics-

Résumé

This paper proposes an automatic procedure, based on ROOT data Analysis Framework, for the analysis of Silicon Photomultipliers (SiPM) characteristics. In particular, it can be used to analyze experimental waveforms, from oscilloscope, containing SiPM pulses acquired at different temperatures and bias voltages. Important SiPMs characteristics such as: charge distribution, gain, breakdown voltage, pulse shape (rise time and recovery time) and overvoltage can been calculated. Developed procedure can be easily used to analyze any type of SiPM detectors.

Dates et versions

hal-02198355 , version 1 (30-07-2019)

Identifiants

Citer

A. Nagai, N. Dinu-Jaeger, A. Para. Silicon Photomultiplier for Medical Imaging -Analysis of SiPM characteristics-. 2019. ⟨hal-02198355⟩
66 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More