Working with the ESEM at high temperature - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Materials Characterization Année : 2019

Working with the ESEM at high temperature

Résumé

Working with the (Environmental) Scanning Electron Microscope (ESEM) at high temperature has been a long lasting goal for many researchers and remains to this day a difficult technique to operate efficiently. This article describes the properties of several different furnaces developed over the past 40 years, focusing specifically on the operation of the high temperature stage associated with the ESEM. Guidelines and advices for correct use of these systems are provided regarding the experience of the authors. A focus on sintering studies illustrates the main difficulties to implement this experimental technique.

Domaines

Matériaux
Fichier principal
Vignette du fichier
S1044580318330523.pdf (14.79 Mo) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-02064126 , version 1 (22-10-2021)

Licence

Paternité - Pas d'utilisation commerciale

Identifiants

Citer

Renaud Podor, G.I. Nkou Bouala, J. Ravaux, J. Lautru, Nicolas Clavier. Working with the ESEM at high temperature. Materials Characterization, 2019, 151, pp.15-26. ⟨10.1016/j.matchar.2019.02.036⟩. ⟨hal-02064126⟩