On the development of a 3D tomographic atom-probe - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Applied Surface Science Année : 1993

On the development of a 3D tomographic atom-probe

Bernard Deconihout
A. Menand
  • Fonction : Auteur
J. M. Sarrau
  • Fonction : Auteur
M. Bouet
  • Fonction : Auteur
S. Chambreland
D. Blavette

Résumé

The development of a three-dimensional atom-probe relies on the design of a position sensitive detector. The multidetection system that we have developed is based on the use of a 10 × 10 anodes array placed behind two microchannel-plates assembly in a chevron arrangement. The spread of charge between MCP and the multianode is used to derive the position of ions striking the detector. Spatial coordinates can be calculated for multiple and simultaneous time-of-flight events. The procedure that we have used for the derivation of ions positions from charge measurement is given and preliminary results are described. \textcopyright 1993.

Dates et versions

hal-01928974 , version 1 (20-11-2018)

Identifiants

Citer

Bernard Deconihout, A. Bostel, A. Menand, J. M. Sarrau, M. Bouet, et al.. On the development of a 3D tomographic atom-probe. Applied Surface Science, 1993, 67 (1-4), pp.444--450. ⟨10.1016/0169-4332(93)90351-B⟩. ⟨hal-01928974⟩