EFFECT OF PATIENT SIZE, ANATOMICAL LOCATION AND MODULATION STRENGTH ON DOSE DELIVERED AND IMAGE-QUALITY ON CT EXAMINATION

Abstract : To study the effect of patient size, anatomical location and modulation strength (MS) on image-quality and delivered dose of CT scans acquired with automatic-exposure-control system (AEC). Four anthropomorphic phantoms (three paediatric and one thin adult) were studied, and normal and obese adults were simulated by placing bolus plates around the adult phantom. Thorax and abdomen-pelvis CT were performed using an AEC system equipped with five possible MS. Modulated tube current (mAsmod) was compared to Reference mAs and image-noise was assessed. Effective-mAs were lower than Reference-mAs for all but the obese phantom. However, reversal points were estimated for an effective diameter of 27.8 cm in thorax and 26.9 cm in abdomen-pelvis scans, beyond which the patterns of MS were inversed. mAsmod were dependent on attenuation differences among distinct anatomical location. Finally, dose delivered was associated to the mAsmod and patient's size, with both affecting image-quality.
Complete list of metadata

https://hal.archives-ouvertes.fr/hal-01895782
Contributor : Anthony Herrada Connect in order to contact the contributor
Submitted on : Monday, October 15, 2018 - 2:56:43 PM
Last modification on : Tuesday, March 15, 2022 - 9:39:02 AM

Identifiers

Collections

Citation

Joël Greffier, Ahmed Larbi, Francesco Macri, Jean-Paul Beregi, F. Pereira. EFFECT OF PATIENT SIZE, ANATOMICAL LOCATION AND MODULATION STRENGTH ON DOSE DELIVERED AND IMAGE-QUALITY ON CT EXAMINATION. Radiation Protection Dosimetry, Oxford University Press (OUP), 2017, 177 (4), pp.373 - 381. ⟨10.1093/rpd/ncx055⟩. ⟨hal-01895782⟩

Share

Metrics

Record views

147