Comparison of Different Ion Mobility Setups Using Poly (Ethylene Oxide) PEO Polymers: Drift Tube, TIMS, and T-Wave

Résumé

Over the years, polymer analyses using ion mobility-mass spectrometry (IM-MS) measurements have been performed on different ion mobility spectrometry (IMS) setups. In order to be able to compare literature data taken on different IM(-MS) instruments, ion heating and ion temperature evaluations have already been explored. Nevertheless, extrapolations to other analytes are difficult and thus straightforward same-sample instrument comparisons seem to be the only reliable way to make sure that the different IM(-MS) setups do not greatly change the gas-phase behavior. We used a large range of degrees of polymerization (DP) of poly(ethylene oxide) PEO homopolymers to measure IMS drift times on three different IM-MS setups: a homemade drift tube (DT), a trapped (TIMS), and a traveling wave (T-Wave) IMS setup. The drift time evolutions were followed for increasing polymer DPs (masses) and charge states, and they are found to be comparable and reproducible on the three instruments.

Dates et versions

hal-01704016 , version 1 (08-02-2018)

Identifiants

Citer

Jean Haler, Philippe Massonnet, Fabien Chirot, Christopher Kune, Clothilde Comby-Zerbino, et al.. Comparison of Different Ion Mobility Setups Using Poly (Ethylene Oxide) PEO Polymers: Drift Tube, TIMS, and T-Wave. Journal of The American Society for Mass Spectrometry, 2018, 29 (1), pp.114-120. ⟨10.1007/s13361-017-1822-9⟩. ⟨hal-01704016⟩
103 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More