Characterization of the aerosol produced by infrared femtosecond laser ablation of polyacrylamide gels for the sensitive inductively coupled plasma mass spectrometry detection of selenoproteins - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Spectrochimica Acta Part B: Atomic Spectroscopy Année : 2009

Characterization of the aerosol produced by infrared femtosecond laser ablation of polyacrylamide gels for the sensitive inductively coupled plasma mass spectrometry detection of selenoproteins

Résumé

A 2D high repetition rate femtosecondlaserablation strategy (2-mm wide lane) previously developed for the detection of selenoproteins in gel electrophoresis by inductively coupled plasma mass spectrometry was found to increase signal sensitivity by a factor of 40 compared to conventional nanosecond ablation (0.12-mm wide lane) [G. Ballihaut, F. Claverie, C. Pécheyran, S. Mounicou, R. Grimaud and R. Lobinski, Sensitive Detection of Selenoproteins in Gel Electrophoresis by High Repetition Rate FemtosecondLaserAblation-Inductively Coupled Plasma Mass Spectrometry, Anal. Chem. 79 (2007) 6874–6880]. Such improvement couldn't be explained solely by the difference of amount of material ablated, and then, was attributed to the aerosol properties. In order to validate this hypothesis, the characterization of the aerosolproduced by nanosecond and high repetition rate femtosecondlaserablation of polyacrylamidegels was investigated. Our 2D high repetition rate femtosecondlaserablation strategy of 2-mm wide lane was found to produce aerosols of similar particle size distribution compared to nanosecond laserablation of 0.12-mm wide lane, with 38% mass of particles < 1 µm. However, at high repetition rate, when the ablated surface was reduced, the particle size distribution was shifted toward thinner particle diameter (up to 77% for a 0.12-mm wide lane at 285 µm depth). Meanwhile, scanning electron microscopy was employed to visualize the morphology of the aerosol. In the case of larger ablation, the fine particles ejected from the sample were found to form agglomerates due to higher ablation rate and then higher collision probability. Additionally, investigations of the plasma temperature changes during the ablation demonstrated that the introduction of such amount of polyacrylamidegel particles had very limited impact on the ICP source (ΔT~ 25 ± 5 K). This suggests that the cohesion forces between the thin particles composing these large aggregates were weak enough to have negligible impact on the ICPMS detection.
Fichier principal
Vignette du fichier
Alexis_6496.pdf (882.92 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-01560756 , version 1 (09-02-2022)

Identifiants

Citer

F. Claverie, Christophe Pécheyran, Sandra Mounicou, G. Ballihaut, B. Fernandez, et al.. Characterization of the aerosol produced by infrared femtosecond laser ablation of polyacrylamide gels for the sensitive inductively coupled plasma mass spectrometry detection of selenoproteins. Spectrochimica Acta Part B: Atomic Spectroscopy, 2009, 64 (7), pp.649-658. ⟨10.1016/j.sab.2009.05.027⟩. ⟨hal-01560756⟩
68 Consultations
18 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More