Laser scanning thermoreflectance imaging system using galvanometric mirrors for temperature measurements of microelectronic devices - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Review of Scientific Instruments Année : 2007

Laser scanning thermoreflectance imaging system using galvanometric mirrors for temperature measurements of microelectronic devices

Résumé

We present a thermoreflectance imaging system using a focused laser sweeping the device under test with a scanner made of galvanometric mirrors. We first show that the spatial resolution of this setup is submicrometric, which makes it adapted to microelectronic thermal measurements. Then, we studied qualitative temperature variations on two dissipative structures constituted of thin (0.35μm) dissipative resistors, the distance between two resistors being equal to 0.8 or 10μm. This technique combines sensitivity and speed: it is faster than a point classical thermoreflectance technique and, in addition, more sensitive than a charge-coupled device thermoreflectance imaging technique.
Fichier non déposé

Dates et versions

hal-01553046 , version 1 (03-07-2017)

Identifiants

Citer

S. Grauby, A. Salhi, Jean-Michel Rampnoux, Hélène Michel, W. Claeys, et al.. Laser scanning thermoreflectance imaging system using galvanometric mirrors for temperature measurements of microelectronic devices. Review of Scientific Instruments, 2007, 78 (7), pp.074902 (1-8). ⟨10.1063/1.2757473⟩. ⟨hal-01553046⟩