Continuous Monitoring of the Vadose Zone Gas Phase by Mass Spectrometry - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Vadose Zone Journal Année : 2016

Continuous Monitoring of the Vadose Zone Gas Phase by Mass Spectrometry

Résumé

Long-term continuous measurements of gas concentrations in the vadose zone are required to identify gas transport processes. A magnetic sector mass spectrometer, originally developed for industrial process monitoring, was deployed in the field, and its performance was assessed for continuous monitoring of vadose zone gases. This instrument allows unattended concentration measurements of a large number of gas species without any purification. With a detection limit on the order of 10 nL L −1 , it allows determination of trace gases such as Kr and Xe isotopes at their natural atmospheric abundances. Good precision and accuracy were obtained for major, minor, and trace gases with a daily calibration. Continuous monitoring of vadose zone gases for 6 mo in two boreholes at various depths at the Roselend Natural Laboratory (French Alps) showed strong spatial and temporal variability of gas concentrations, not only O 2 and CO 2 but also inert gases. Such time series allow determination of transport processes and interactions between water and gas fluxes in the vadose zone. Abbreviations: FC, Faraday cup; SEM, secondary electron multiplier.
Fichier non déposé

Dates et versions

hal-01483371 , version 1 (06-03-2017)

Identifiants

Citer

Sophie Guillon, C Gréau, E Pili. Continuous Monitoring of the Vadose Zone Gas Phase by Mass Spectrometry. Vadose Zone Journal, 2016, 15, pp.0 - 0. ⟨10.2136/vzj2015.12.0168⟩. ⟨hal-01483371⟩
136 Consultations
1 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More