Determination of the isotopic composition of micrometric uranium particles by UV femtosecond laser ablation coupled with sector-field single-collector ICP-MS. - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Journal of Analytical Atomic Spectrometry Année : 2017

Determination of the isotopic composition of micrometric uranium particles by UV femtosecond laser ablation coupled with sector-field single-collector ICP-MS.

Résumé

The coupling of a UV-fs-LA system with a sector-field ICP-MS for isotopic anal. of individual sub micrometric and micrometric uranium particles, esp. for the measurement of minor isotopes (234U and 236U), is discussed. Ablation of uranium particles results in short (a few seconds) and highly noisy signals with sudden and extremely brief (\textless1 ms) increases in intensity (referred to as 'spikes'), making the precise detn. of isotope ratios very challenging. Uranium particles were located beforehand by suitable techniques. Three isotope ratio calcn. methods are evaluated. The influence of signal spikes on the accuracy and precision of the isotope ratios is discussed. Thanks to the implementation of a 'spike rejection filter', the measurement of uranium isotope ratios in particles from an isotopically certified ref. material (NBS U010) is greatly improved. Measured ratios are in acceptable agreement with the ref. values although a systematic bias was obsd. Limits of detection of a few tens of attograms for minor isotopes are achieved. Isotopic mapping in the area around the initial ablation crater allows the observation of the deposition of debris. No isotope ratio bias is obsd. when analyzing uranium particles mixed with lead particles, showing that no significant lead-based polyat. interference is produced. Results, including minor isotope ratios, obtained with particles of sizes from 0.3 to 1.5 μm sampled in a nuclear facility, are in good agreement with the values obtained by another well-established method for such anal.
Fichier non déposé

Dates et versions

hal-01481299 , version 1 (02-03-2017)

Identifiants

Citer

Ariane Donard, F. Pointurier, A.-C. Pottin, A. Hubert, Christophe Pécheyran. Determination of the isotopic composition of micrometric uranium particles by UV femtosecond laser ablation coupled with sector-field single-collector ICP-MS.. Journal of Analytical Atomic Spectrometry, 2017, 32, pp.96-106. ⟨10.1039/C6JA00071A⟩. ⟨hal-01481299⟩
56 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More