Low-noise cold-field emission current obtained between two opposed carbon cone nanotips during in situ transmission electron microscope biasing

Abstract : A dedicated transmission electron microscope sample holder has been used to study in situ the cold-field emission process of carbon cone nanotips (CCnTs). We show that when using a CCnT instead of a Au plate-anode, the standard deviation of the emission current noise can be decreased from the 10 nA range to the 1 nA range under vacuum conditions of 10−5 Pa. This shows the strong influence of the anode on the cold-field emission current noise.
Type de document :
Article dans une revue
Applied Physics Letters, American Institute of Physics, 2015, 106 (26), pp.263101. 〈10.1063/1.4923245〉
Liste complète des métadonnées

Littérature citée [13 références]  Voir  Masquer  Télécharger

https://hal.archives-ouvertes.fr/hal-01430586
Contributeur : Martin Hytch <>
Soumis le : vendredi 23 février 2018 - 14:05:47
Dernière modification le : mercredi 18 avril 2018 - 16:36:33
Document(s) archivé(s) le : vendredi 25 mai 2018 - 04:00:03

Fichier

1.4923245.pdf
Fichiers éditeurs autorisés sur une archive ouverte

Identifiants

Collections

Citation

Ludvig De Knoop, Christophe Gatel, Florent Houdellier, Marc Monthioux, Aurélien Masseboeuf, et al.. Low-noise cold-field emission current obtained between two opposed carbon cone nanotips during in situ transmission electron microscope biasing. Applied Physics Letters, American Institute of Physics, 2015, 106 (26), pp.263101. 〈10.1063/1.4923245〉. 〈hal-01430586v2〉