Uncertainty of reference frames characterized by real time optical measurements: Application to Computer Assisted Orthopaedic Surgery - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue CIRP Annals - Manufacturing Technology Année : 2009

Uncertainty of reference frames characterized by real time optical measurements: Application to Computer Assisted Orthopaedic Surgery

Résumé

In total knee arthroplasty (TKA), real time optical measurements are now carried out to improve the implantation of prostheses. The final alignment of prostheses is however influenced by the choice of points probed by the surgeon, and by the measurement accuracy. A Monte Carlo method has been developed, to simulate a whole set of measurements. To estimate the uncertainties of the resultant angles between the two parts of the knee prosthesis, this data has then been treated with the algorithms implemented in Computer Assisted Orthopaedic Surgery programs. Such results were also obtained through uncertainty propagations based on the GUM.
Fichier principal
Vignette du fichier
Linares_16_uncertainly_of_reference_december2009.pdf (630.31 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-01426919 , version 1 (05-01-2017)

Identifiants

Citer

Jean-Marc Linares, Jean-Michel Sprauel, P. Bourdet. Uncertainty of reference frames characterized by real time optical measurements: Application to Computer Assisted Orthopaedic Surgery. CIRP Annals - Manufacturing Technology, 2009, 58 (1), pp.447 - 450. ⟨10.1016/j.cirp.2009.03.008⟩. ⟨hal-01426919⟩
33 Consultations
118 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More