Electrochemically prepared polypyrrole-2-carboxylic acid films: synthesis protocols and studies on biosensors

Abstract : The process for obtaining polypyrrole-2-carboxylic acid (PPY-2-COOH) films in acetonitrile was investigated using cyclic voltammetry, electrochemical quartz crystal microgravimetry (EQCM), and infrared spectroscopy (FTIR). Different potential ranges were applied during cyclic voltammetry experiments with the aim of obtaining films without and with the presence of controlled amounts of water added in acetonitrile. The FTIR spectra of the films have evidenced that cations and anions from the electrolyte solution were incorporated into the PPY-2-COOH structure, with a preferential adsorption of cations. After chemically immobilizing polyphenoloxidase (tyrosinase, PPO), PPY-2-COOH/PPO films were build for amperometric detection of catechol, establishing a linear limit of concentrations ranging from 5.0x104 to 2.5x102molL1.
Document type :
Journal articles
Liste complète des métadonnées

https://hal.archives-ouvertes.fr/hal-00878377
Contributor : Agnès Bussy <>
Submitted on : Wednesday, October 30, 2013 - 8:56:17 AM
Last modification on : Thursday, February 8, 2018 - 11:10:19 AM

Links full text

Identifiers

Collections

Citation

M. Foschini, A. Marletta, R. C. Faria, Didier Léonard, François Bessueille, et al.. Electrochemically prepared polypyrrole-2-carboxylic acid films: synthesis protocols and studies on biosensors. Electroanalysis, Wiley-VCH Verlag, 2013, 25 (3), pp.741-749. ⟨10.1002/elan.201200574⟩. ⟨hal-00878377⟩

Share

Metrics

Record views

110