Influence of Geant4 parameters on dose distribution and computation time for carbon ion therapy simulation.

Abstract : The aim of this work was to study the influence of Geant4 parameters on dose distribution and computational time for simulations of carbon ion therapy. The study was done using Geant4 version 9.0. The dose distribution in water for incident monoenergetic carbon ion beams of 300 MeV/u were compared for different values of secondary particle production threshold and different step limits. Variations of depth dose of about 2 mm were observed in some cases, which induced a 30% variation of dose deposit in the Bragg peak region. Other tests were done using Geant4 version 9.2 to verify the results from this study. The two versions provided converging results and led to the same conclusions.
Document type :
Journal articles
Physica Medica, Elsevier, 2010, 26 (4), pp.202-8. <10.1016/j.ejmp.2009.12.001>


https://hal.archives-ouvertes.fr/hal-00633360
Contributor : Pascal Calvat <>
Submitted on : Tuesday, October 18, 2011 - 1:15:19 PM
Last modification on : Monday, October 13, 2014 - 3:43:25 PM

Identifiers

Collections

Citation

N. Zahra, T. Frisson, L. Grevillot, P. Lautesse, D. Sarrut. Influence of Geant4 parameters on dose distribution and computation time for carbon ion therapy simulation.. Physica Medica, Elsevier, 2010, 26 (4), pp.202-8. <10.1016/j.ejmp.2009.12.001>. <hal-00633360>

Export

Share

Metrics

Consultation de la notice

32