Stability criterions of an oscillating tip-cantilever system in dynamic force microscopy - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue The European Physical Journal B: Condensed Matter and Complex Systems Année : 2001

Stability criterions of an oscillating tip-cantilever system in dynamic force microscopy

Résumé

This work is a theoretical investigation of the stability of the non-linear behavior of an oscillating tip-cantilever system used in dynamic force microscopy. Stability criterions are derived that may help to a better understanding of the instabilities that may appear in the dynamic modes, Tapping and NC-AFM, when the tip is close to a surface. A variational principle allows to get the temporal dependance of the equations of motion of the oscillator as a function of the non-linear coupling term. These equations are the basis for the analysis of the stability. One find that the branch associated to frequencies larger than the resonance is always stable whereas the branch associated to frequencies smaller than the resonance exhibits two stable domains and one unstable. This feature allows to re-interpret the instabilities appearing in Tapping mode and may help to understand the reason why the NC-AFM mode is stable.
Fichier principal
Vignette du fichier
EPJB.pdf (356.81 Ko) Télécharger le fichier
Loading...

Dates et versions

hal-00004544 , version 1 (22-03-2005)

Identifiants

Citer

Laurent Nony, Rodolphe Boisgard, Jean-Pierre Aimé. Stability criterions of an oscillating tip-cantilever system in dynamic force microscopy. The European Physical Journal B: Condensed Matter and Complex Systems, 2001, 24, pp.221-229. ⟨hal-00004544⟩
67 Consultations
129 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More