Doppler Radar for Heartbeat Rate and Heart Rate Variability Extraction
Dany Obeid ( ) 1, Sawsan Sadek 2, Gheorghe Zaharia 3, Ghaïs El Zein 3
CEDRE, AUF Collaboration(s)
(24/11/2011)

This paper presents a Doppler radar system used to detect the heartbeat signal from a d√istance of one meter. The proposed system is based on using a vector network analyzer and two antennas. Measurements are performed at 16 GHz for different power levels between 0 and -25 dBm. Both heartbeat rate and heart rate variability are extracted and compared to a simultaneous ECG signal.
1 :  Institut d'Electronique et de Télécommunications de Rennes (IETR)
CNRS : UMR6164 – Université de Rennes 1 – Institut National des Sciences Appliquées (INSA) - Rennes – SUPELEC – Université de Nantes
2 :  IUT de Saida
Université Libanaise
3 :  Institut d'Electronique et de Télécommunications de Rennes (IETR)
CNRS : UMR6164 – Université de Rennes 1 – Institut National des Sciences Appliquées (INSA) - Rennes – SUPELEC
Systèmes de communications numériques (SCN)
Propagation, localisation (PL)
Sciences de l'ingénieur/Electronique
microwave systems * Doppler effect * noncontact detection * cardiopulmonary activity * heartbeat rate