Development and Validation of a Potentiometric Biosensor Assay for Tylosin with Demonstrated Applicability for Detection of Two Other Antimicrobial Growth Promoter Compounds in Feedstuffs - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue Food Additives and Contaminants Année : 2011

Development and Validation of a Potentiometric Biosensor Assay for Tylosin with Demonstrated Applicability for Detection of Two Other Antimicrobial Growth Promoter Compounds in Feedstuffs

Sara Lucy Stead
  • Fonction : Auteur correspondant
  • PersonId : 896488

Connectez-vous pour contacter l'auteur
Katarzyna Wolodko-Cierniak
  • Fonction : Auteur
  • PersonId : 925952
Sarah Fiona Richmond
  • Fonction : Auteur
  • PersonId : 925953
Matthew Sharman
  • Fonction : Auteur
  • PersonId : 896443
Olga Leonardova
  • Fonction : Auteur
  • PersonId : 925954
Duncan Purvis
  • Fonction : Auteur
  • PersonId : 925955
Phil Driver
  • Fonction : Auteur
  • PersonId : 925956
Phil Teale
  • Fonction : Auteur
  • PersonId : 917589

Résumé

A potentiometric biosensor assay based on a commercially available polyclonal antibody has been developed to detect tylosin residues in animal feed. The method can be used as a rapid (less than 45 minutes) laboratory-based procedure or, as a portable field-test for the simultaneous measurement of up to 12 different samples. For both procedures the qualitative detection capability (CC) for tylosin was determined as 0.2 mg kg-1 in a range of animal feeds with a measurement repeatability at concentrations between 0.2 and 4 mg kg-1 of ≤13% %Coefficient of Variation (%CV). The field-test format was capable of detecting tylosin residues at operating (external air) temperatures ranging between +4 to 37oC although some reduction in signal was observed at the lower temperatures. The laboratory-based tylosin assay was evaluated using 16 medicated and 22 non-medicated feeds and found to give comparable data to a confirmatory method based upon Liquid Chromatography tandem Mass Spectrometry (LC-MS/MS). The potential to develop a multi-probe format assay for the simultaneous detection of tylosin, spiramycin and virginiamycin was also demonstrated. Cross-validation in a second laboratory showed the assay to be transferable, reliable and robust.

Mots clés

Fichier principal
Vignette du fichier
PEER_stage2_10.1080%2F19440049.2011.567399.pdf (564.76 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-00704664 , version 1 (06-06-2012)

Identifiants

Citer

Sara Lucy Stead, Katarzyna Wolodko-Cierniak, Sarah Fiona Richmond, Matthew Sharman, Olga Leonardova, et al.. Development and Validation of a Potentiometric Biosensor Assay for Tylosin with Demonstrated Applicability for Detection of Two Other Antimicrobial Growth Promoter Compounds in Feedstuffs. Food Additives and Contaminants, 2011, 28 (07), pp.848-859. ⟨10.1080/19440049.2011.567399⟩. ⟨hal-00704664⟩

Collections

PEER
99 Consultations
145 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More