Transmission line excited chamber : experimental and numerical analysis
Résumé
This paper presents experimental and numerical results concerning Transmission Line Excited Chamber (TLEC). This chamber is an alternative for electromagnetic compatibility tests, notably radiated immunity. The E-field profile, frequency response, SWR, and S parameters are addressed and compared. Besides application regarding performance assessment of electromedical equipment, this chamber could also be applied as a way to reload intra-body devices or to get biotelemetry information.
Domaines
Energie électrique
Origine : Fichiers produits par l'(les) auteur(s)